Skip to content

    Navigation breadcrumbs

  1. Home
  2. Veterinary topics and resources
  3. All resources
  4. Comparison between a flash glucose monitoring system and a portable blood glucose meter for monitoring dogs with diabetes mellitus

Library and information services

Access to electronic and print resources focused on veterinary science and animal health and services to support your study and keep up to date with clinical research.

Awards and prizes

Our awards celebrate achievements and build knowledge that contributes to evidence-based veterinary medicine.

History

We hold a unique collection of books, archives, artefacts and memorabilia which together offer an insight into the evolution of the British veterinary profession.

    Navigation breadcrumbs

  1. Home
  2. Veterinary topics and resources
  3. All resources
  4. Comparison between a flash glucose monitoring system and a portable blood glucose meter for monitoring dogs with diabetes mellitus
Journal watch24 February 2021

Comparison between a flash glucose monitoring system and a portable blood glucose meter for monitoring dogs with diabetes mellitus

Evidence-based veterinary medicineSmall animalsMedicine

Author(s): F. Del Baldo, C. Canton, S. Testa, H. Swales, I. Drudi, S. Golinelli and F. Fracassi
Published in: Journal of Veterinary Internal Medicine
Date: November/December 2020
DOI: https://doi.org/10.1111/jvim.15930
Type of access: Open Access
(click for full article)

Our summary

Del Baldo, F. et al. (2020) Comparison between a flash glucose monitoring system and a portable blood glucose meter for monitoring dogs with diabetes mellitus. Journal of Veterinary Internal Medicine, 34 (6), pp. 2296-2305.

The aim of this prospective observational study was to compare glucose curve (GC) readings taken from insulin-treated dogs with diabetes using two different monitoring systems: The Freestyle Libre Flash Glucose Monitoring System (FGMS) and the Portable Blood Glucose Meter (PBGM). Investigators also analysed the data collected to compare the accuracy of the two systems in recording the GC nadir, and to characterise GCs obtained both in different environments (hospital vs home) and at different times (nocturnal vs diurnal).

The study was carried out at a single university veterinary hospital between May 2015 and March 2018. Twenty dogs with diabetes mellitus (DM), who had been treated with insulin for at least four weeks prior to recruitment, were enrolled in the study. All dogs were fitted with a FGMS and interstitial glucose curves (IGCs) were acquired for each dog during the 14-day recording period. On study days 1, 7 and 14, paired in-hospital GCs for each dog were acquired using the FGMS and PBGM.  The remaining four IGCs were obtained at home by the owner using the FGMS.

Summary glucose reports (the Ambulatory Glucose Profile (AGP)) generated by the FGMS included a daily log showing interstitial glucose fluctuations between 20 and 350 mg/dL during a 24-hour period, this was used to provide an estimation of glucose nadirs and the number of low glucose episodes. Hypothetical insulin dose recommendations were then made for each GC, and comparisons were made for the in-hospital GCs, from the FGMS and PBGM, for the consecutive day IGCs acquired at home and for the consecutive day home vs hospital IGCs acquired with the FGMS.

Results showed that when comparing GCs acquired in hospital, using both the FGMS and the PBCM, the insulin dosing recommendation would have been the same in 33/43 cases and 34/43 cases considering the glycaemic nadir. For IGCs performed at home on two consecutive days using the FGMS, the insulin dosing recommendation would have only been the same in 5/14 cases and 9/14 cases when considering the glycaemic nadir. For IGCs performed on two consecutive days at home and hospital using the FGMS, the insulin dosing recommendation would have been the same in 17/25 cases and 16/25 for glycaemic nadir.

When assessing nadir, the AGP was used as a ‘gold standard’. Using the AGP, glucose nadirs were identified in 34/43 (79%) IGCs. When considering the 34 IGCs in which a glucose nadir was identified, concordant glucose nadirs were detected in 28/34 (82%) IGCs generated by the individual FGMS scans and in 14/34 (41%) BGCs obtained by PBGM.

Limitations of the study include the small number of dogs enrolled and that, for this study, FGMS was considered the gold standard. Additionally, for 12/20 dogs, the sensor in the FGMS did not last for the 14-day period, which reduced the amount of data available for analysis and meant that the accuracy of the FGMSs were not evaluated for the entire study period.

Take home

This study shows that the FGMS could be a valuable tool in detecting glucose nadirs in dogs with diabetes mellitus and that the real-time data collected and the reports generated by the AGP software allow for retrospective analysis of day-to-day variations in glycaemic control. The study highlights that there are limitations in both forms of monitoring.

If you found this article interesting you may also be interested in
Deiting, V. and Mische, R. (2021) Use of the “FreeStyle Libre” glucose monitoring system in diabetic cats. Research in Veterinary Science, 135, pp. 253-259. https://doi.org/10.1016/j.rvsc.2020.09.015.

Next steps

Receive journal watch by email

Subscribe to have the latest summaries sent to your inbox

Claim CPD credit for your reading

Reading and reflecting on articles can count towards your CPD, and we have a template to help you with the process